Abstract
Nowadays, lateral flow immunochromatographic assays are increasingly popular as a diagnostic tool for point-of-care (POC) test based on their simplicity, specificity, and sensitivity. Hence, quantitative detection and pluralistic popular application are urgently needed in medical examination. In this study, a smartphone-based dual-modality imaging system was developed for quantitative detection of color or fluorescent lateral flow test strips, which can be operated anywhere at any time. In this system, the white and ultra-violet (UV) light of optical device was designed, which was tunable with different strips, and the Sobel operator algorithm was used in the software, which could enhance the identification ability to recognize the test area from the background boundary information. Moreover, this technology based on extraction of the components from RGB format (red, green, and blue) of color strips or only red format of the fluorescent strips can obviously improve the high-signal intensity and sensitivity. Fifty samples were used to evaluate the accuracy of this system, and the ideal detection limit was calculated separately from detection of human chorionic gonadotropin (HCG) and carcinoembryonic antigen (CEA). The results indicated that smartphone-controlled dual-modality imaging system could provide various POC diagnoses, which becomes a potential technology for developing the next-generation of portable system in the near future.
Background
In recent decades, lateral flow immunochromatographic strip (LFICS) has been increasingly applied as a diagnostic tool for point-of-care test (POCT) because of its simplicity, rapid speed, sensitivity, and specificity [1]. Nowadays, this technology has been applied for many areas such as food security, environment monitoring, and drug testing [2–8]. In order to meet the requirement of different detection, diverse signal marker strips had been developed, including colloidal gold strip, magnetic nanoparticles strip, and fluorescence strip [9–12]. As the typical nanomaterials, gold nanoparticles (Au NPs) have the color-tunable property with size, which also can be functionalized with peptides to avoid degradation in order to facilitate to finish quick immunological reaction and display different degree of color for half-quantitative detection [1, 11, 13]. Because of their unique advantages, such as simplicity, rapidity, and ease of interpretation, Au NPs-based LFICS was widely used for rapid diagnosis [1, 11, 13]. Magnetic nanoparticles (MNPs) could be easily gathered through directing external magnetic field exposure; thus, the targeted molecule modified on the MNPs could be detected even with less amount used [14]. Magnetic signals could be entirely captured by the devices so that MNPs-based LFICS realized quantitative measurement, thereby increasing LFICS sensitivity [12, 14, 15]. Fluorescent NPs possessed many advantages, including photo-stability, wide absorption and narrow emission spectra, more noticeable size-dependent Stokes shifts, robust stability against photo-bleaching and chemical degradation, higher fluorescent quantum yield, high sensitivity, and large molar extinction coefficients [16–18]. These excellent properties rendered fluorescent NPs playing an important role for developing highly sensitive LFICS. Different kinds of fluorescence LFICS were studied, such as testing chloramphenicol, nitrated ceruloplasmin, ochratoxin, alpha fetoprotein [16–20].
Until now, the typical detection based on LFICS is through the naked eye, which was widely used in food safety, environment monitoring, and precision medicine. However, this detection can only provide qualitative test (positive or negative) or semi-quantitative information on analyte concentration, which the LFICS could not satisfy requirements for practical applications. Therefore, many devices providing quantitative analyte concentration for testing LFICS had been developed [9, 21]. Some strip readers were designed to work in the desktop computer or laptop, which possessed rapid processing speeds and stable performances. However, these bulky and heavy devices limited their wide application for the trend of family and personalize care. Mei’s group developed an embedded system based on the Acorn RISC Machine (ARM) processor, which applied for reading a test strip [22]. Marquina’s group designed a spin-valve giant magnetoresistive (GMR) sensor system to quantify the amount of analyte, which were not portable for outdoor detection [23]. Obviously, the strip reader based on mobile device could be satisfied with the requirement of high portability and feature-rich testing. As a matter fact, the mobile health market is rapidly developing and portable diagnostics tools provide an opportunity to increase the availability of healthcare and decrease costs [24]. Therefore, various strip readers based on mobile device have been developed. For example, Oncescu’s group developed a smartphone accessory and software application based on smartphone that allowed for the quantification of cholesterol levels in the blood [25] and Hyun Park’s group developed a kind of strip reader for quantitative measurements of H5N1 using fluorescent strips [26]. Meanwhile, some other strip readers based on smartphone were obtained, such as testing thyroid-stimulating hormone, cholesterol, malaria, and tuberculosis [9, 27–32]. However, all the strip readers based on smartphone only can detect one type of strips from the color or fluorescence which limited their multifunctional applications [33–36].
In this study, smartphone-based dual-modality imaging system was developed, which could quantitatively detect color or fluorescent lateral flow immunochromatographic strip (ICTS). In this system, the white and UV light of an optical system was designed and could be changed according to the different kinds of strips (color strip or fluorescence strip). The improved Sobel operator was used in the software, which improved greatly the ability of distinguishing between the test area and background boundary information. This smartphone provided the power to the whole system and could be continuously operated for 4–5 h.
The detection limit of the reader reached the requirements as the limit values were 2.3 mIU/mL and 0.037 ng/mL separately, which were calculated separately from detection of color strip (human chorionic gonadotropin, HCG) and fluorescence strip (carcinoembryonic antigen, CEA). To our knowledge, it is one kind of novel dual-modality imaging for immunochromatographic strip based on smartphone and has great potential in food security, environment monitoring, and drug testing in near future.
Methods
Composition of Test Strips
The antigen and antibody of the HCG and CEA were both purchased from the Shanghai Linc-Bio Science Co. LTD (shanghai, China). Conjugate, sample, and absorbent pads as well as the nitrocellulose (NC) membrane and polyvinyl chloride (PVC) plate were purchased from JieYi Biotech Co., Ltd. (Shanghai, China). Immunochromatographic strips were prepared by our group’s member. The test strip was composed of a sample pad, conjugate pad, and absorbent pad, as well as a nitrocellulose (NC) membrane and PVC backing card, as shown in Fig. 1. All pretreated parts were assembled sequentially onto a PVC backing card with 2 mm overlap of each component. The assembly was cut into 3-mm-wide individual strips and then stored at 4 °C inside a sealed plastic bucket with a desiccant until further use.
Primary antibody (Ab2) against the antigen was immobilized on the test line (T line), and another antibody (Ab1) against a different epitope of the same antigen was labeled with NPs. When the NP-labeled antibody flowed along with the test sample (which was expected to have the antigen), it interacted with the antigen in the sample. This complex interacted with the antibody coated at the test line on the membrane and formed a sandwich (Fig. 1). A control line was fixed with the second antibody against the primary antibody, which captured the remaining antibody labeled with NPs. Thus, in a sandwich assay format, a positive test sample is indicated by the appearance of two colored lines (T line and C line) and a negative test sample was indicated by the appearance of one colored line that is only the control line (C line).
Design of Hardware
Dual-modality imaging system was developed based on smartphone. In order to decrease the cost and increase the cost performance, Red Mi Note 2 smart phone (Xiaomi mobile internet company) was chosen as the main part. The smartphone was equipped with Central Processing Unit (CPU, MediaTek Helio X10 MT6795 processor) and Graphics Processing Unit (GPU, PowerVR G6200), and its processing speed and computing performance were greatly for image handling. Even though this smartphone screen has a high resolution (1920 × 1080 p) and the screen pixel density is 441 ppi. The rear camera is based on complementary metal-oxide-semiconductor (CMOS) image sensor with 13 million pixels. Therefore, using smartphones as a data acquisition, image processing, and light module was appropriate. Black smartphone accessory (55 × 78 × 30 mm3) and an ICTS cartridge (70 × 18 × 4.8 mm3) were designed by SolidWorks software and fabricated using a 3D printer (Fig. 2a). The attachment module included optical path structure, lens, filter, sample slot, LED, and some fixed structure. Two ultra-violet (UV) LEDs (365 nm, 1 W) were fixed on both sides of the upper end of the cartridge, and the flash of smartphone was selected as the white light. The filter and lens group has been fixed in the accessory box. Appropriate design of optical paths (Fig. 2b) not only provided homogeneous optic field but also shortened the focusing distance, which helped us get the high-quality images within just 20-mm height shell. Dual-modality imaging (using white LED or UV LED) can be switched easily according to different kinds of strip (colloidal gold strip and fluorescence strips). The smartphone provided power to the whole system and could be operated continuously in 4–5 h.
Software Features
A complete dual-modality imaging system required not only hardware but also powerful and easily operated software. The Java programming language was employed in the design of Android Studio 1.5 development environment and provided main functions of processing test strip images, analysis, and diagnosis in the smartphone. The system has the features of accuracy, friendly and mostly. The dual-modality imaging system not only can test the colloidal gold strips through the colorimetry but also can test the quantum dot strips through the florescence signal.
The function of software included the detection and analysis of cloud data, cloud data storage, cloud data queries, intelligent robot service consultant, and location of the user. After the strip is inserted, the user could choose the detection mode (UV or white LED) to open the camera and click the test button, and then get the analysis results while waiting for 3 s. All the result data would be stored in the cloud and could be queried when the user needed them. Through the system, the user could get the geographical distribution of the patients by analysis the cloud data. In the system, the intelligent robot consultant was also developed, which could provide some useful messages according to the user’s questions, see in Fig. 3a. In the flow diagram of the software system in Fig. 3b, the orange color part was used for the non-registered patients, and the test results still could be saved in our cloud data manager system. The green flow chart is used for the registered users. Once registered, the users can get much more permission to access the software function, such as managing their own detection results and getting the advices from the intelligent robot consultant.
In Fig. 3b, it showed that saving and inquiring data were the based function which was for upholding the software, and then, the big data analysis and the intelligent advisor was the advance function. Through the data analysis, the user distribution was easily viewed on the map, which can provide government disease control. And through the intelligent advisor, the users could get some useful advices according their own healthy condition. With the rapid development of big data and cloud storage, clinical monitoring data contains a great value. In order to take advantage of excavation clinical data value, we developed a corresponding cloud data management system which emerged in the dual-modality imaging system. The system could achieve high-volume statistical data analysis, query, and management and could be classified according to the statistics geographical distribution of the disease, which would provide a reference for the epidemiological investigation.
Clinical Specimen Examination
This study was approved by the Medical Ethics Committee of Shanghai Jiao Tong University, China. All patients signed an informed consent, and all methods were performed in accordance with the relevant guidelines and regulations. All the clinical samples were collected from Shanghai Ninth People’s Hospital, the affiliated hospital of Shanghai Jiao Tong University. Total 50 urea-enzymes samples and 50 antigens of CEA were tested using a dual-modality imaging system employing lateral flow strips.
Data Statistic
All data are presented in this paper as mean ± standard deviation. Statistical differences were evaluated using the t test and considered significant at P < 0.05.
Results and Discussion
To produce accurate diagnosis, a high-resolution image was acquired. The camera in the smartphone was based on CMOS image sensor with 13 million pixels. It has excellent optical character, high purity, whiteness, high resolution, and superb color rendition. In the dual-modality imaging system, the camera could be operated conveniently by clicking the control button. In order to obtain more accurate signal data, multiple image processing was applied, such as gray-scale processing, improved Sobel convolution operator, threshold analysis, image binarization, and getting detection area boundary coordinates, as shown in Fig. 4. All the images were captured every 0.1 s and three times in total to obtain the average value.
Research of image edge examination
Once the image data was obtained, the signal area in the image was then identified. Compared with many other algorithms [6, 28–31], this algorithm could be easily be used in smartphone. Converting the three-channel image into a single-channel grayscale image mainly simplify the calculation of subsequent processing on gray-scale process. In this case, it is very convenient to find out the feature boundary of the detecting area. Sobel convolution operator algorithm was proposed to process the image and advantaged to search the boundary of T and C lines. The procedure of the proposed algorithm is listed below.
-
1.
Sx is the x direction calculation result. Sy is the y direction calculation result. S(x, y) is the result of convolution operation after the f(x, y) function.
$$ S x=\left[3 f\left( x+1,\ y\mathit{\hbox{-}}1\right)+10 f\left( x+1, y\right)+3 f\left( x+1, y+1\right)\right]\mathit{\hbox{-}}\left[3 f\left( x\mathit{\hbox{-}}1,\ y\mathit{\hbox{-}}1\right)+10 f\left( x\mathit{\hbox{-}}1,\ y\right)+3 f\left( x\mathit{\hbox{-}}1,\ y+1\right)\right] $$(1)$$ S y=\left[3 f\left( x\mathit{\hbox{-}}1,\ y\mathit{\hbox{-}}1\right)+10 f\left( x, y\mathit{\hbox{-}}1\right)+3 f\left( x+1, y\mathit{\hbox{-}}1\right)\right]\mathit{\hbox{-}}\left[3 f\left( x\mathit{\hbox{-}}1,\ y+1\right)+10 f\left( x,\ y+1\right)+3 f\left( x+1,\ y+1\right)\right] $$(2)$$ S\left( x, y\right)=\sqrt{\left( Sx\hat{\mkern6mu} 2+ Sy\hat{\mkern6mu} 2\right)} $$(3) -
2.
Matrix is shown below in (4), which was Sobel mask operator. The matrix was used in Eqs. (1) and (2).
For the boundary line of the acquisition, the use of Sobel operator was much better in comparison with that of other edge detection operators.
Besides, even though there are many advantages for linear detection with use the typical way of Hough line detection algorithm, the strip mainly composes of straight lines but software needs much more time for using this algorithm, which limited its use in this technology. Figure 5a showed the time-consuming algorithm comparison.
The next step was to select the gray image threshold before it was binary processing. It could acquire a much better accurate threshold value after combining the optimal threshold segmentation principles and experience value. The effect of the optimal threshold method alone was acceptable, yet it was also a more time-consuming algorithm. To avoid that drawback, it was a fast and simple way to combine the algorithm and the reference value being stored in the database to calculate the threshold. After setting 0 (black) to the pixel value above the threshold and setting 255 (white) to the pixel below above the threshold, binarization image was easily obtained. From Fig. 5b, we can see that there are too much signal edge noise points after the image was processed by the Sobel operator and Prewitt operator, with wide boundary signal region. Besides, there is no obvious boundary from both sides of the image after treated with the Roberts operator. Meanwhile, the image noise points were still not decreased after the process of the Log operator, while the one which treated with the improved operator can clearly see the left and right boundaries of the strip, with clear and narrow signal region boundary. Therefore, after comparison of these five algorithms (improved Sobel convolution operator algorithm, Sobel algorithm, Prewitt algorithm, Log algorithm, and Roberts algorithm), apparently, the improved Sobel convolution operator algorithm was the best one which derived a clear boundary and had less noise point.
The following step was to get the boundary coordinate of T and C lines through the binarization image. By calculating the left and right limits of test area, the coordinate in the Y-axis of the T/C line could be calibrated by two horizontal lines as shown in Fig. 5c. To confirm the accurate coordinate value, center expansion method (points of ic and id in Fig. 5c) and extreme method (points of ia and ib in Fig. 5c) were compared, which showed the former algorithm was much more effective. More noise signal existed beyond the area of the ia and ib. The procedure of the proposed algorithm is listed below.
-
1.
The binarization image was the M × N matrix. Calculating the sum of every column pixel (5) in binarization image as the function X(i), the function X(i) was displayed in Fig. 5c.
$$ X(i)={\displaystyle {\sum}_{j=0}^M} S\left({x}_i,{y}_j\right) $$(5)Obtaining the two extreme points (points of ia and ib in Fig. 5c) through Eqs. (6) and (7).
$$ {X}_{\mathrm{extr}1}=\begin{array}{c}\hfill N/2\hfill \\ {}\hfill i=0\hfill \end{array}\mathrm{Max}\left( X(i)\right)= X\left(\mathrm{ia}\right) $$(6)$$ {X}_{\mathrm{extr}2}=\kern1.44em \begin{array}{c}\hfill N\hfill \\ {}\hfill i= N/2\hfill \end{array}\mathrm{Max}\left( X(i)\right)= X\left(\mathrm{ib}\right) $$(7)After then, the half of the smaller extreme value (points X(ia) and X(ib) in Fig. 5c) was set as the new threshold value through Eq. (8).
$$ T=\frac{1}{2}\mathrm{M}\mathrm{i}\mathrm{n}\left({X}_{\mathrm{etr}1}+{X}_{\mathrm{extr}2}\right) $$(8) -
2.
From the middle to both sides in horizontal axis, the value in the X(i) was compared with the threshold. Once the value was bigger than the threshold, the corresponding points (ic, id) were established through Eqs. (9) and (10).
Through the extreme method, the two extreme points (ia and ib) of the coordinate position are acquired. Meanwhile, the center expansion method was used to calculate and to get much more accuracy value (ic and id) of (9) and (10).
As we know, T and C lines were two rectangular areas. By obtaining the left and right edge coordinates (points of ic and id in Fig. 5c) of the T/C line, the efficiency of the detection area was identified. The next work was to calculate the other two sides, as shown in Fig. 5d.
-
1.
According to the varying degrees of definition in images, two methods were used. The first one is directly summing each row of the pixel values to obtain a one-dimensional array aim at obvious T/C lines.
$$ Y(j)={\displaystyle {\sum}_{i=0}^N\; S\left({x}_i,{y}_j\right)} $$(11) -
2.
Through calculating four extreme values (12–17), the extreme points (c1, c2, t1, t2) corresponded to the boundary coordinate of T and C lines.
-
3.
Aiming at unobvious T and C lines, the second method was to contrast the four extreme values (c1, c2, t1, and t2) and statistic of the averaging boundary coordinates of the T/C line in the previous experiments and then revised them.
Combining the two methods, the effective area of the test strip could be automatically identified, which was a great progress compared to other detection devices [32].
Extraction for Detection Image Signals
According to different strips labeled by different NPs, we designed two methods to extract the signal of the strips: color strip and fluorescent strips. In the color strip, the pixel value of the red, green, and blue (RGB) three channels was extracted as the strip signal in T line and C line of the original image because every channel of the RGB three channels has indicated the intensity of the signal. In the fluorescent strips, only the red channel was extracted as the fluorescent intensity signal. From Fig. 6, we can see the green and blue channels could not indicate the fluorescent intensity and even had a negative effect on the signal extraction. So, the pixel value of the red channel was appropriate as the fluorescent intensity signal.
Detection of Clinical Samples
The system was used to test the clinical samples. HCG and CEA samples (30 positive and 20 negative) were collected from a hospital. All the samples were pretreated with the treating fluid which was prepared by our group’s members. A double-blind test was used in this experiment. In the clinical samples, electrochemiluminescence assay was used to determined human HCG and commercial ECLIA kit (Roche Cobase 601) was the reference for CEA test. If one truly positive sample in the clinical samples was negatively detected by our method, the result was defined as false negative, and vice versa. The detection results were shown in Table 1. The sensitivity and specificity of the device were 97 and 100% for HCG and were 97 and 95% for CEA, respectively. From Table 1, it is demonstrated that the detecting system had a high sensitivity and specificity.
Accuracy of the System
In order to confirm the system accuracy, different bio-reagent concentration of the test strips was applied and the results were shown in Fig. 7. Figure 7a was the colloidal gold strips detection for HCG with six concentrations of analyte standards (0, 10, 15, 30, 60, and 120 mIU/mL) and Fig. 7b was the fluorescence strips detection for CEA with eight concentration of standard antigen (1, 2.5, 5, 10, 20, 30, 40, and 50 ng/mL). All the samples were prepared in a standard dilution buffer. Different concentration of the strips were detected for 20 times. From the results, we can see the areas of C and T lines were automatically marked correctly by the red circle even the vague shape of T line. Extracting the signal of the tagged area in the strips and calculating the ratio of T and C could further improve the sensitivity and the specificity of the system.
After recording the signal, the standard curve was obtained by plotting the linearity of T/C against the concentration of HCG or CEA, as represented by the equation in Fig. 7d, f. With an increase in concentration, more antigens were captured in the T line, leading to an increase in the T/C ratio. Meanwhile, the linear correlation coefficient (R 2) values were relatively high, and the ideal detection limit of the system was found to be 2.3 mIU/mL for HCG and 0.037 ng/mL for CEA.
Repeatability and Stability of the System
The different light intensities and strip locations caused slight variations in the acquired image. To test the repeatability of the system, the same concentration of ten strips was detected repeatedly and every strip was inserted into the system for 10 times within 5 min. In order to evaluate the stability of the system, immunochromatographic strips containing three different concentrations (10, 60, and 120 mIU/mL for HCG and 5, 20, and 50 ng/mL for CEA) were tested by the system. This test was performed to detect the error tolerance of the variation, and the results were presented in Fig. 8. The variance was calculated by the results after tested for 10 times at each strip. For HCG, 10, 60, 120 mIU/mL concentrations corresponding to the detection of the standard deviation were respectively 2.63, 1.37, 0.91%. For CEA, 5, 20, 50 ng/mL concentrations corresponding to the detection of the standard deviation were respectively 3.85, 1.45, and 1.05%. The variances of the whole test results were 1.6% (colored strip) and 2.1% (fluorescent strip), respectively. From the results, we can conclude that stability of the system is proportional to the sample concentration, and it presented the similar values from each measurement, thus proving the good repeatability of the system.
Conclusions
In summary, a novel smartphone-based dual-modality imaging system was developed, which could quantitatively detect color or fluorescent lateral flow immunochromatographic strip (ICTS). Reasonable algorithms were applied to calculate the boundary localization and improve the accuracy of extracting signal, which achieved the high-signal intensity and sensitivity. The performance of the system was tested using different samples, which presented satisfactory results. Since there are many connected objects designed to link up with smartphone-controlled system and big data to realize bio-analysts and personalized healthcare monitoring and management, further work will focus on the development of a portable and versatile imaging system based on smartphone for detection in POCT in the near future.
References
Wang K, Qin W, Hou Y, Xiao K (2016) The application of lateral flow immunoassay in point of care testing: a review. Nano Biomed Eng 8:172–183
Liu L, Luo L, Suryoprabowo S, Peng J, Kuang H, Xu C (2014) Development of an immunochromatographic strip test for rapid detection of ciprofloxacin in milk samples. Sensors 14:16785–16798
Chen Y, Kong D, Liu L, Song S, Kuang H, Xu C (2015) Development of an ELISA and immunochromatographic assay for tetracycline, oxytetracycline, and chlortetracycline residues in milk and honey based on the class-specific monoclonal antibody. Food Anal Methods 9:905–914
Gopinath SC, Tang TH, Chen Y, Citartan M, Lakshmipriya T (2014) Bacterial detection: from microscope to smartphone. Biosens Bioelectron 60:332–342
Singh J, Sharma S, Nara S (2015) Evaluation of gold nanoparticle based lateral flow assays for diagnosis of enterobacteriaceae members in food and water. Food Chem 170:470–483
Ren M, Xu H, Huang X, Kuang M, Xiong Y, Xu H et al (2014) Immunochromatographic assay for ultrasensitive detection of aflatoxin B(1) in maize by highly luminescent quantum dot beads. ACS Appl Mater Interfaces 6:14215–14222
Yetisen AK, Martinez-Hurtado JL, Garcia-Melendrez A, Vasconcellos F, Lowe CR (2014) A smartphone algorithm with inter-phone repeatability for the analysis of colorimetric tests. Sensors Actuators B Chem 196:156–160
Yang H, Deng M, Ga S, Chen SH, Kang L et al (2014) Capillary-driven surface-enhanced Raman scattering (SERS)-based microfluidic chip for abrin detection. Nanoscale Res Lett 9:138
Qin W, Wang K, Xiao K et al (2017) Carcinoembryonic antigen detection with “Handing”-controlled fluorescence spectroscopy using a color matrix for point-of-care applications. Biosens Bioelectron 90:508–515
Yan X, Wang K, Lu W, Qin W, Cui D, He J (2016) CdSe/ZnS quantum dot-labeled lateral flow strips for rapid and quantitative detection of gastric cancer carbohydrate antigen 72–4. Nanoscale Res Lett 11:138
Huang X, Aguilar Z, Xu H, Lai W, Xiong Y (2016) Membrane-based lateral flow immunochromatographic strip with nanoparticles as reporters for detection: a review. Biosens Bioelectron 75:166–180
Zhang XQ, Jiang L, Zhang CL, Li D et al (2011) A silicon dioxide modified magnetic nanoparticles-labeled lateral flow strips for HBs antigen. J Biomed Nanotechnol 7:776–781
Chen Y, Cui D (2013) Non-spherical gold nanoparticles: tumor imaging and therapy. Nano Biomed Eng 5:160–167
Chen Z, Wen J, Ju H, Fang Z (2015) Magnetic nano-Fe3O4 particles targeted gathering and bio-effects on nude mice loading human hepatoma Bel-7402 cell lines model under external magnetic field exposure in vivo. Electromagn Biol Med 34:309–316
Chen Y, Wang K, Liu Z, Sun R, Cui D, He J (2016) Rapid detection and quantification of tumor marker carbohydrate antigen 72–4 (CA72-4) using a superparamagnetic immunochromatographic strip. Anal Bioanal Chem 408:2319–2327
Wang L, Chen W, Ma W, Liu L, Ma W, Zhao Y et al (2011) Fluorescent strip sensor for rapid determination of toxins. Chem Commun 47:1574–1576
Duan H, Chen X, Xu W, Fu J, Xiong Y, Wang A (2015) Quantum-dot submicrobead-based immunochromatographic assay for quantitative and sensitive detection of zearalenone. Talanta 132:126–131
Taranova N, Berlina A, Zherdev A, Dzantiev B (2015) Traffic light’ immunochromatographic test based on multicolor quantum dots for the simultaneous detection of several antibiotics in milk. Biosens Bioelectron 63:255–261
Yang Q, Gong X, Song T, Yang J, Zhu S, Li Y et al (2011) Quantum dot-based immunochromatography test strip for rapid, quantitative and sensitive detection of alpha fetoprotein. Biosens Bioelectron 30:145–150
Liang R, Xu X, Liu T, Zhou J, Wang X, Ren Z et al (2015) Rapid and sensitive lateral flow immunoassay method for determining alpha fetoprotein in serum using europium (III) chelate microparticles-based lateral flow test strips. Anal Chim Acta 891:277–283
Oncescu V, Mancuso M, Erickson D (2014) Cholesterol testing on a smartphone. Lab On A Chip 14:759–763
Guo T, Patnaik R, Kuhlmann K, Rai A, Sia S (2015) Smartphone dongle for simultaneous measurement of hemoglobin concentration and detection of HIV antibodies. Lab On A Chip 15:3514–3520
Mei J, Ye Q, Zhou W, Hui W, Tian J (2011) development and study of lateral flow test strip reader based on embedded system. The Tenth International Conference on Electronic Measurement & Instruments 201–204
Marquina C, Teresa J, Serrate D, Marzo J, Cardoso F, Cardoso S et al (2012) GMR sensors and magnetic nanoparticles for immuno-chromatographic assays. J Magn Magn Mater 324:3495–3498
Oncescu V, O’Dell D, Erickson D (2013) Smartphone based health accessory for colorimetric detection of biomarkers in sweat and saliva. Lab Chip 13:3232–3238
Feng S, Caire R, Cortazar B, Turan M, Wong A, Czcan A (2014) Immunochromatographic diagnostic test analysis using Google glass. ACS Nano 8:3069–3079
You D, Park T, Yoon J (2013) Cell-phone-based measurement of TSH using Mie scatter optimized lateral flow assays. Biosens Bioelectron 40:180–185
Mudanyali O, Dimitrov S, Sikora U, Padmanabhan S, Navruz I, Ozcan A (2012) Integrated rapid-diagnostic-test reader platform on a cellphone. Lab on A Chip 12:2678–2686
Yeo S, Huong D, Hong N, Li C, Choi K, Yu K et al (2014) Rapid and quantitative detection of zoonotic influenza A virus infection utilizing coumarin-derived dendrimer-based fluorescent immunochromatographic strip test (FICT). Theranostics 4:1239–1249
Zheng Y, Wang K, Zhang J, Qin W, Yan X, Shen G et al (2016) Simultaneous quantitative detection of Helicobacter pylori based on a rapid and sensitive testing platform using quantum dots-labeled immunochromatiographic test strips. Nanoscale Res Lett 11:62
Zangheri M, Cevenini L, Anfossi L, Baggiani C, Simoni P, Nardo F et al (2015) A simple and compact smartphone accessory for quantitative chemiluminescence-based lateral flow immunoassay for salivary cortisol detection. Biosens Bioelectron 64:63–68
Ozkan H, Kayhan O (2016) A novel automatic rapid diagnostic test reader platform. Comput Math Methods Med 74:98217
Shi X, Liu H, Song Y (2015) Pollutional haze as a potential cause of lung cancer. J Thorac Dis 7:E412–E417
Huang L, Zhou L, Zhang Y, Xie C, Qu J, Zeng A et al (2009) A simple optical reader for upconverting phosphor particles captured on lateral flow strip. IEEE Sensors J 9:1185–1191
Qin Z, Chan W, Boulware D, Akkin T, Butler E, Bischof J (2012) Significantly improved analytical sensitivity of lateral flow immunoassays by using thermal contrast. Angew Chem 51:4358–4361
Gui C, Wang K, Li C, Dai X, Cui D (2014) A CCD-based reader combined with CdS quantum dot-labeled lateral flow strips for ultrasensitive quantitative detection of CagA. Nanoscale Res Lett 9:57
Acknowledgements
We are grateful for the financial support by the National Natural Scientific Foundation of China (Grant Nos. 81571835, 61503246, 81672247, and 81671737), National Key Basic Research Program (973 Project) (No. 2015CB931802), the 863 High-Tech Project of China (No. 2014AA020700), and Shanghai Science and Technology Fund (No. 13NM1401500 and 15DZ2252000).
Authors’ Contributions
YH jointly conceived the study with KW. KX and WL performed the experiments. YH, WQ, JN, and JS performed all the data analysis and wrote the original draft of the manuscript. WT and DC reviewed and edited the manuscript. All authors reviewed the manuscript. All authors read and approved the final manuscript.
Authors’ Information
K.W. is an associate professor of Shanghai Jiao Tong University. Her scientific interests are nanotechnology development of early cancer detection and screening equipment, nonmaterial molecular imaging, and biocompatibility evaluation. D.C. is a professor of Shanghai Jiao Tong University. His research interests include the synthesis of nanomaterials and their application in the biomedical field. W.T. is an associate professor of Shanghai Jiao Tong University. Her research interests include biosensors, photoelectric detection, and laser inspection. K.X. and W.Q. are both master students of Shanghai Jiao Tong University. W.L. is a master student of Zhujiang Hospital, Southern Medical University.
Competing Interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Author information
Authors and Affiliations
Corresponding authors
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Hou, Y., Wang, K., Xiao, K. et al. Smartphone-Based Dual-Modality Imaging System for Quantitative Detection of Color or Fluorescent Lateral Flow Immunochromatographic Strips. Nanoscale Res Lett 12, 291 (2017). https://doi.org/10.1186/s11671-017-2078-9