Table 2 Comparison of analytical characteristics of biosensors based on urease adsorbed on silicalite-1-covered transducers

From: Nanosized zeolites as a perspective material for conductometric biosensors creation

Biosensor characteristic

Biosensors based on transducers covered via spin coating

Biosensors based on transducers covered via drop coating

Time of steady-state response

1 min

1.5 to 2 min

Sensitivity to urea, μS/mM

700 to 860

1,100 to 1,500

Linear range, μM

10 to 1,000

2 to 700

Limit of urea detection, μM

5 to 10

1 to 2

Reproducibility of responses during working day (RSD)

3% to 8%

2% to 4%

Reproducibility of biosensor preparation (RSD of responses of different biosensors)

12% to 15%

20% to 25%