Table 4 Recovery of V. cholerae J3324 and V. cholerae UVC1324 DNAs by using the proposed PSA-AuNPs-based electrochemical DNA biosensor

From: Ultrasensitive Biosensor for the Detection of Vibrio cholerae DNA with Polystyrene-co-acrylic Acid Composite Nanospheres

Bacteria (μg/μL)

Current (μA), n = 3

Found (μg/μL)

Recovery (%)

Vibrio cholerae J3324

 1.0 × 10−4

12.82 ± 0.35

1.07 × 10−4

107.2

 1.0 × 10−5

11.56 ± 0.22

9.14 × 10−6

91.4

 1.0 × 10−6

10.41 ± 0.37

9.55 × 10−7

95.5

Vibrio cholerae UVC1324

1.0 × 10−4

12.76 ± 0.56

9.57 × 10−5

95.7

1.0 × 10−5

11.61 ± 0.43

1.01 × 10−6

100.5

1.0 × 10−6

10.48 ± 0.48

1.09 × 10−6

108.9