Table 1 The reproducibility measurement of the blue Mini-LEDs MOD

From: Using Blue Mini-LEDs as a Light Source Designed a Miniaturized Optomechanical Device for the Detection of Direct Bilirubin

Direct bilirubin concentration (μmol/L)

The absorbance for the first test (arb. u)

The absorbance for a second test (arb. u)

The absorbance for the third test (arb. u)

The absorbance for the fourth test (arb. u)

The absorbance for the fifth test (arb. u)

Standard deviation error (%)

1.71

0.680

0.674

0.674

0.673

0.672

0.140

8.55

0.354

0.352

0.351

0.351

0.350

0.068

17.1

0.090

0.089

0.089

0.087

0.086

0.073