Table 1 Comparison of the parameters of the used analysis methods

From: Analysis methods for meso- and macroporous silicon etching baths

Parameter

Titration

F-ISE

Cuvette test

Analyte

HF

HF

Organic content

Measurement method

Indirect

Direct

Direct

Measurement range

>50 mmol L−1

0.05 to 50 mmol L−1

0.08 to 2.7 mmol L−1

Time for measurement

15 min

5 min

4 h

Time for calibration

5 mina

15 minb

Only once

Suitable for acetic acid containing electrolytes?

No

Yes

Yes

Relative error

3.3 %

1.5 %

2.1 %

Standard deviation

0.1 mol L−1(n = 10)

0.2 mol L−1(n = 10)

0.01 mol L−1(n = 20)

Inline analysis possible?

Yes

Yes

No

  1. aOnce a week. bDirectly before measurement.