Skip to main content
Account

Table 5 A comparison of the developed DNA biosensor performance with other previously reported electrochemical DNA biosensor for the determination of pathogenic V. cholere DNA

From: Ultrasensitive Biosensor for the Detection of Vibrio cholerae DNA with Polystyrene-co-acrylic Acid Composite Nanospheres

Material and electrode design

Linear range (M)

LOD (M)

Hybridi-sation time (min)

Reference

DNA-PSA-AuNPs-carbon SPE (Direct immobilisation of capture probe on PSA gold-matrix)

1.0 × 10−21–1.0 × 10−8

1.0 × 10−21

60

Present work

AuNPs-PSA-Avidin conjugate label on reporter probe

1.0 × 10−18–1.0 × 10−15

1.0 × 10−15

30

Liew et al. (2015) [24]

AuNP-latex label on DNA reporter probe (Avidin-biotin binding)

1.0 × 10−15–1.0 × 10−12

1.0 × 10−16

-

Kuan at el. (2013) [30]

AuNP-latex label on DNA reporter probe (Avidin-biotin binding)

1.0 × 10−12–1.0 × 10−9

0.5 × 10−16

20

Pinijsuwan et al. (2008) [29]

AuNP-latex label on DNA target (streptavidin-latex/biotin-AuNP)

1.0 × 10−12–1.0 × 10−9

1.0 × 10−12

20

Kawde & Wang (2003) [35]