Table 2 Dark current and output current measured before and after lifetime testing

From: High-Sensitivity and Long-Life Microchannel Plate Processed by Atomic Layer Deposition

 

Dark current before lifetime testing (pA)

Dark current after lifetime testing (pA)

Output current before lifetime testing (pA)

Output current after lifetime testing (pA)

Traditional MCP

1.0

6.5

1300

630

ALD-MCP

1.2

1.5

5700

6000